You Will Receive:
– Thermo Scientific Shandon CryoStar NX50 HOVPD Cryostat Microtome (957130)
– Knife/Blade Holder Assembly
– x3 Specimen Discs
– Heat Extractor
– CryoStar NX70/NX50 Operator Guide
– Power Cable
– 90-Day Warranty
HOVPD Option Configuration Includes:
Option H – Height adjustment that raises and lowers the work area
Option O – Specimen cooling
Option V – Vacutome
Option P – Peltier cooling cryobar
Option D – Cold disinfection
Fully Tested and Working Without Issue:
After its initial disinfection and inspection, our technicians tested the Thermo Scientific Shandon CryoStar NX50 HOVPD cryostat microtome with the following results:
– Specimen cooling -40°C setpoint (manufacturer’s spec): PASSED
– Height adjustment: PASSED
– Peltier cooling: PASSED
– Vacutome: PASSED
Notable Service Performed:
– Inspected refrigeration lines.
– Flushed and filled tyfoxit fluid for specimen cooling.
– Tested nebulizer for disinfectant option.
Specifications:
Manufacturer | Thermo Fisher Scientific |
Model | CryoStar NX50 HOVPD |
Ref. No. | 957130 |
Serial No. | S20010233 |
Fine section thickness range | 0.5 – 100 µm |
Trimming thickness range | 5 – 500 µm |
Vertical specimen stroke | 64 mm (2.5 in) |
Horizontal knife movement | 48 mm (1.9 in) |
Specimen orientation | x- and y- axes universal 8° with zero positioning; z-axis up to 360° |
Specimen retraction | 20 µm |
Specimen temperature control | -43°C ±2°C (-45.4°F) at 20°C (68°F) ambient |
Cryobar cooling | -57°C ±3°C (-70.6°F) at 20°C (68°F) ambient |
Cryochamber | -25°C ±2°C (-13°F) at 20°C (68°F) ambient; continuously cooled |
Knife holder | -27°C ±2°C (-16.6°F) at 20°C to 30°C (68°F to 86°F) ambient; continuously cooled |
Defrosting options | Programmed / Immediate |
Certifications | CSA, CE |
Power rating | 110-120V~, 60Hz, 10A |
Dimensions (W x D x H) | 33 x 38 x 43 to 54.5 in |
Weight | 441 lb (200 kg) |